Soft-db Opus Suite Data Logger Module User Manual Page 57

  • Download
  • Add to my manuals
  • Print
  • Page
    / 62
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 56
56
Appendix 2: 1/3 Octave Filters – IEC 61260 Class 1/ANSI S1.11
1/3 Octave Filters
The 1/3 octave filters are computed at low-level in real time (at 48 kHz) on the digital signal
processor (DSP) of the Concerto system. The filters comply with all requirements of IEC 61260 for
Class 1.
Frequency Range
20 Hz to 20 kHz.
Filter Shape
The following curve presents the filter shape test done for the 1000 Hz 1/3 octave band. The red and
green curves represent the limits associated with the IEC standard (Class 1).
1/3 octave Filter Shape Test at 1000 Hz
-100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
100 1000 10000
Frequence (Hz)
Attenuation dB
Measurement High limit Low limit